Time-of-Flight measurements with a detector using a liquid Cherenkov radiator-prototype of a possible TOF detector for the Super-FRS at FAIR

N. Kuzminchuk-Feuerstein, O. Bogdanov, E. Rozhkova, C. Scheidenberger, B. Voss

Результат исследований: Материалы для журналаСтатьярецензирование

2 Цитирования (Scopus)

Аннотация

Aiming to develop a Time-of-Flight (TOF) detector with a precision down to about 50 ps (σ) in time, radiation hard and operating with a high radiation rate of relativistic heavy ions of up to 107 per spill of 1 to 10 s., we have constructed a new Time-of-Flight Cherenkov detector based on an Iodine-Naphthalene liquid radiator. This fluid with high refractive index (n=1.7003) was chosen to cover the corresponding velocities of the ions β∼ 0.59-0.92 at the future Super-FRS at FAIR. The application of a liquid radiator allows to circulate the material and therefore greatly reduce the effects of a degradation of the optical performance during exposure to the high ion rates and radiation dose at the focal planes of the Super-FRS. The key properties of the prototype of a TOF-Cherenkov detector have been investigated in measurements with nickel ions at GSI in 2014. More recently, TOF measurements were performed with xenon ions at 600 MeV/u at GSI in 2016 and a timing resolution of 63 ps (standard deviation σ) was achieved.

Язык оригиналаАнглийский
Страницы (с-по)34-37
Число страниц4
ЖурналNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Том923
DOI
СостояниеОпубликовано - 11 апр 2019

ASJC Scopus subject areas

  • Nuclear and High Energy Physics
  • Instrumentation

Fingerprint Подробные сведения о темах исследования «Time-of-Flight measurements with a detector using a liquid Cherenkov radiator-prototype of a possible TOF detector for the Super-FRS at FAIR». Вместе они формируют уникальный семантический отпечаток (fingerprint).