The effect of the regime of recording diffractograms on the profile and width of instrumental lines

S. I. Kuznetsov, A. S. Panin, E. Yu Tarasova

Результат исследований: Материалы для журналаСтатья

Аннотация

The influence of the operating mode of an X-ray diffractometer on the profile and integral width of instrumental lines was investigated. It was shown that the choice of a function for the best approximation of the instrumental line depends on an angular range of recording. The best approxi- mation functions in different angular ranges were determined. It was shown that the main effect on the instrumental line width is exerted by the dimensions of the input vertical divergence and receiving slits. The criterion of choice of the maximum size of the receiving slit for X-ray tubes with different sizes of the focal spots was experimentally substantiated.

Язык оригиналаАнглийский
Страницы (с-по)299-308
Число страниц10
ЖурналRussian Journal of Nondestructive Testing
Том48
Номер выпуска5
DOI
СостояниеОпубликовано - мая 2012

ASJC Scopus subject areas

  • Mechanical Engineering
  • Mechanics of Materials
  • Condensed Matter Physics
  • Materials Science(all)

Fingerprint Подробные сведения о темах исследования «The effect of the regime of recording diffractograms on the profile and width of instrumental lines». Вместе они формируют уникальный семантический отпечаток (fingerprint).