Release of Ni from NiTi substrates covered with a protective calcium phosphate coating deposited by rf-magnetron sputtering

Результат исследований: Материалы для книги/типы отчетовМатериалы для конференции

1 Цитирования (Scopus)

Аннотация

According to the data presented in Figure 2, the average Ni release rate from an uncoated NiTi substrate was 7.3 ng cm -2 d -1 and for a coated NiTi substrate 0.8 ng cm -2 d -1. Therefore, the Ni release from a coated sample was 9 times smaller than that from an uncoated one. Different preparation techniques for the calcium phosphate coating (plasma atmosphere, Figure 2) did not have a significant influence. We therefore conclude that the main reason is that all coatings are continuous, well adhered and pore-free. We tentatively ascribe the small Ni release from coated NiTi samples to the fact that their edges were not fully covered with a protective coating.

Язык оригиналаАнглийский
Название основной публикации8th World Biomaterials Congress 2008
Страницы1759
Число страниц1
Том4
СостояниеОпубликовано - 2008
Событие8th World Biomaterials Congress 2008, WBC 2008 - Amsterdam, Нидерланды
Продолжительность: 28 мая 20081 июн 2008

Другое

Другое8th World Biomaterials Congress 2008, WBC 2008
СтранаНидерланды
ГородAmsterdam
Период28.5.081.6.08

ASJC Scopus subject areas

  • Biomaterials

Fingerprint Подробные сведения о темах исследования «Release of Ni from NiTi substrates covered with a protective calcium phosphate coating deposited by rf-magnetron sputtering». Вместе они формируют уникальный семантический отпечаток (fingerprint).