In this paper the numerical simulation results of the dose spatial distribution of the medical electron beams in ABS-plastic doped with different concentrations of lead and zinc are shown. The dependences of the test material density on the lead and zinc mass concentrations are illustrated. The depth dose distributions of the medical electron beams in the modified ABS-plastic for three energies 6 MeV, 12 MeV and 20 MeV are tested. The electron beam shapes in the transverse plane in ABS-plastic doped with different concentrations of lead and zinc are presented.