Determination of 1-Hydroxypyrene Using High-Performance Liquid Chromatography with Electrochemical Detection

J. Barek, V. Bencko, J. Cvačka, V. Mejstřík, A. Slámová, I. Švagrová, J. Zima

Результат исследований: Материалы для журналаСтатья

1 Цитирования (Scopus)

Аннотация

Optimum conditions are described for HPLC determination of 1-hydroxypyrene with electrochemical detection (ED) giving the detection limit ∼ 1.10-8 mol.l-1. This detection limit was compared with that for UV spectrophotometric detection (∼ 1.10-7 mol.l-1) and fluorimetric detection (∼ 5.10-10 mol.l-1). It was demonstrated that HPLC-ED can be used for the determination of 1-hydroxypyrene in urine down to 1.10-7 mol.l-1 and thus for biological monitoring of human exposure to polycyclic aromatic hydrocarbons.

Язык оригиналаАнглийский
Страницы (с-по)875-876
Число страниц2
ЖурналChemicke Listy
Том91
Номер выпуска10
СостояниеОпубликовано - 1 дек 1997
Опубликовано для внешнего пользованияДа

ASJC Scopus subject areas

  • Chemistry(all)