A simple approach of presampled modulation transfer function measurement tested on the phoenix nanotom scanner

D. Ivashkov, A. Batranin, T. Mamyrbayev

Результат исследований: Материалы для журнала

Аннотация

In this paper presampled modulation transfer function of the 2D images obtained on the Phoenix Nanotom scanner was investigated with different measurement set-ups. Three parameters were chosen to investigate their influence on modulation transfer function: source- detector distance, tube current and binning mode. A simple method for modulation transfer function determination of digital imaging detectors from edge images was applied. The following results were achieved and briefly discussed: modulation transfer function improves with increase of the source-detector distance, slightly improves with increase of the current and remains constant for different binning modes. All measurements were carried out in University of Applied Sciences Upper Austria at Wels campus.

Язык оригиналаАнглийский
Номер статьи012021
ЖурналJournal of Physics: Conference Series
Том671
Номер выпуска1
DOI
СостояниеОпубликовано - 18 янв 2016
Событие3rd All-Russian Scientific and Practical Conference on Innovations in Non-Destructive Testing, SibTest 2015 - Altai, Российская Федерация
Продолжительность: 27 июл 201531 июл 2015

ASJC Scopus subject areas

  • Physics and Astronomy(all)

Fingerprint Подробные сведения о темах исследования «A simple approach of presampled modulation transfer function measurement tested on the phoenix nanotom scanner». Вместе они формируют уникальный семантический отпечаток (fingerprint).

  • Цитировать