Voltammetric determination of cholesterol in human blood serum

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

Cholesterol has a number of important functions in a human body. The disorders of cholesterol biosynthesis increase the risk of the development of cardiovascular diseases and worsens the function of the immune system. This study is devoted to the development of a method for determining the concentration of cholesterol in blood serum using voltammetry. We selected the following working conditions for the determination of cholesterol: a phosphate buffer solution with pH 6.86 was a supporting electrolyte; potential sweep rate was W = 30 mV/s; and recording was carried out in the constant-current mode. An organomodified electrode was used as a working electrode. The peak was observed at +0.98 V. The dependence of the electrooxidation current of cholesterol on its concentration was linear in the range of 0.1–50 μM with a limit of detection of 0.01 μM. The results of the determination of cholesterol in blood serum by voltammetry were compared to those obtained by the fluorimetric method.

Original languageEnglish
Pages (from-to)904-910
Number of pages7
JournalJournal of Analytical Chemistry
Volume72
Issue number8
DOIs
Publication statusPublished - 1 Aug 2017

Fingerprint

Blood
Cholesterol
Voltammetry
Electrodes
Electrooxidation
Immune system
Biosynthesis
Electrolytes
Buffers
Phosphates

Keywords

  • blood serum
  • cardiovascular diseases
  • cholesterol
  • modifier
  • voltammetry

ASJC Scopus subject areas

  • Analytical Chemistry

Cite this

Voltammetric determination of cholesterol in human blood serum. / Derina, K. V.; Korotkova, E. I.; Dorozhko, E. V.; Voronova, O. A.

In: Journal of Analytical Chemistry, Vol. 72, No. 8, 01.08.2017, p. 904-910.

Research output: Contribution to journalArticle

@article{bcd8b26bec22404eb407db56191c0736,
title = "Voltammetric determination of cholesterol in human blood serum",
abstract = "Cholesterol has a number of important functions in a human body. The disorders of cholesterol biosynthesis increase the risk of the development of cardiovascular diseases and worsens the function of the immune system. This study is devoted to the development of a method for determining the concentration of cholesterol in blood serum using voltammetry. We selected the following working conditions for the determination of cholesterol: a phosphate buffer solution with pH 6.86 was a supporting electrolyte; potential sweep rate was W = 30 mV/s; and recording was carried out in the constant-current mode. An organomodified electrode was used as a working electrode. The peak was observed at +0.98 V. The dependence of the electrooxidation current of cholesterol on its concentration was linear in the range of 0.1–50 μM with a limit of detection of 0.01 μM. The results of the determination of cholesterol in blood serum by voltammetry were compared to those obtained by the fluorimetric method.",
keywords = "blood serum, cardiovascular diseases, cholesterol, modifier, voltammetry",
author = "Derina, {K. V.} and Korotkova, {E. I.} and Dorozhko, {E. V.} and Voronova, {O. A.}",
year = "2017",
month = "8",
day = "1",
doi = "10.1134/S1061934817080068",
language = "English",
volume = "72",
pages = "904--910",
journal = "Journal of Analytical Chemistry",
issn = "1061-9348",
publisher = "Maik Nauka-Interperiodica Publishing",
number = "8",

}

TY - JOUR

T1 - Voltammetric determination of cholesterol in human blood serum

AU - Derina, K. V.

AU - Korotkova, E. I.

AU - Dorozhko, E. V.

AU - Voronova, O. A.

PY - 2017/8/1

Y1 - 2017/8/1

N2 - Cholesterol has a number of important functions in a human body. The disorders of cholesterol biosynthesis increase the risk of the development of cardiovascular diseases and worsens the function of the immune system. This study is devoted to the development of a method for determining the concentration of cholesterol in blood serum using voltammetry. We selected the following working conditions for the determination of cholesterol: a phosphate buffer solution with pH 6.86 was a supporting electrolyte; potential sweep rate was W = 30 mV/s; and recording was carried out in the constant-current mode. An organomodified electrode was used as a working electrode. The peak was observed at +0.98 V. The dependence of the electrooxidation current of cholesterol on its concentration was linear in the range of 0.1–50 μM with a limit of detection of 0.01 μM. The results of the determination of cholesterol in blood serum by voltammetry were compared to those obtained by the fluorimetric method.

AB - Cholesterol has a number of important functions in a human body. The disorders of cholesterol biosynthesis increase the risk of the development of cardiovascular diseases and worsens the function of the immune system. This study is devoted to the development of a method for determining the concentration of cholesterol in blood serum using voltammetry. We selected the following working conditions for the determination of cholesterol: a phosphate buffer solution with pH 6.86 was a supporting electrolyte; potential sweep rate was W = 30 mV/s; and recording was carried out in the constant-current mode. An organomodified electrode was used as a working electrode. The peak was observed at +0.98 V. The dependence of the electrooxidation current of cholesterol on its concentration was linear in the range of 0.1–50 μM with a limit of detection of 0.01 μM. The results of the determination of cholesterol in blood serum by voltammetry were compared to those obtained by the fluorimetric method.

KW - blood serum

KW - cardiovascular diseases

KW - cholesterol

KW - modifier

KW - voltammetry

UR - http://www.scopus.com/inward/record.url?scp=85026678144&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85026678144&partnerID=8YFLogxK

U2 - 10.1134/S1061934817080068

DO - 10.1134/S1061934817080068

M3 - Article

VL - 72

SP - 904

EP - 910

JO - Journal of Analytical Chemistry

JF - Journal of Analytical Chemistry

SN - 1061-9348

IS - 8

ER -