The radiation dose determination of the pulsed X-ray source

Research output: Chapter in Book/Report/Conference proceedingConference contribution

1 Citation (Scopus)

Abstract

In this paper the radiation dose measurement technique of the pulsed X-ray source RAP-160-5 is described. The dose rate measurement results from the pulsed X-ray beams at the different distance between the pulsed X-ray source focus and the detector obtained with the help of the thermoluminescent detectors DTL-02, the universal dosimeter UNIDOS E equipped with the plane-parallel ionization chamber type 23342, the dosimeter-radiometer DKS-96 and the radiation dosimeter AT 1123 are demonstrated. The recommendations for the dosimetry measurements of the pulsed X-ray generator RAP-160-5 under different radiation conditions are proposed.

Original languageEnglish
Title of host publicationIOP Conference Series: Materials Science and Engineering
PublisherInstitute of Physics Publishing
Volume66
Edition1
DOIs
Publication statusPublished - 2014
Event20th International Conference for Students and Young Scientists: Modern Techniques and Technologies, MTT 2014 - Tomsk, Russian Federation
Duration: 14 Apr 201418 Apr 2014

Other

Other20th International Conference for Students and Young Scientists: Modern Techniques and Technologies, MTT 2014
CountryRussian Federation
CityTomsk
Period14.4.1418.4.14

    Fingerprint

ASJC Scopus subject areas

  • Engineering(all)
  • Materials Science(all)

Cite this

Miloichikova, I. A., Stuchebrov, SG., Zhaksybayeva, G., & Wagner, A. (2014). The radiation dose determination of the pulsed X-ray source. In IOP Conference Series: Materials Science and Engineering (1 ed., Vol. 66). [012031] Institute of Physics Publishing. https://doi.org/10.1088/1757-899X/66/1/012031