The construction of the fast resistive bolometer for a SXR measurement on the GIT-12 facility

Jakub Cikhardt, Daniel Klír, Pavel Kubeš, Josef Kravárik, Karel Řezáč, Ondřej Šíla, Alexander V. Shishlov, Alexey Yu Labetsky

Research output: Contribution to journalArticle

Abstract

A lot of kinds of instruments are used for the SXR measurement at pulsed power facilities, but most of them are difficult to calibrate absolutely. For the determination of the energy of SXR radiated by the discharge on Z-pinches, it is possible to use the bolometer which can be calibrated analytically. The bolometer can be constructed with the sufficient sensitivity and, at the same time, with the time resolution in the order of nanoseconds. This bolometer was designed and constructed for the measurement on the 5MA facility GIT-12 at the Institute of High Current Electronics (IHCE) of the Siberian Branch Russian Academy of Sciences in Tomsk. The experiments on GIT-12 with the neon and deuterium gas-puff load were diagnosed by the copper bolometer with the time resolution of 4 ns and the sensitivity of 12Vcm2 J-1.

Original languageEnglish
Pages (from-to)134-137
Number of pages4
JournalActa Polytechnica
Volume53
Issue number2
Publication statusPublished - 2013
Externally publishedYes

Fingerprint

Bolometers
Neon
Deuterium
Electronic equipment
Copper
Gases
Experiments

Keywords

  • Bolometers
  • Plasma diagnostics
  • X-rays
  • Z-pinches

ASJC Scopus subject areas

  • Engineering(all)

Cite this

Cikhardt, J., Klír, D., Kubeš, P., Kravárik, J., Řezáč, K., Šíla, O., ... Labetsky, A. Y. (2013). The construction of the fast resistive bolometer for a SXR measurement on the GIT-12 facility. Acta Polytechnica, 53(2), 134-137.

The construction of the fast resistive bolometer for a SXR measurement on the GIT-12 facility. / Cikhardt, Jakub; Klír, Daniel; Kubeš, Pavel; Kravárik, Josef; Řezáč, Karel; Šíla, Ondřej; Shishlov, Alexander V.; Labetsky, Alexey Yu.

In: Acta Polytechnica, Vol. 53, No. 2, 2013, p. 134-137.

Research output: Contribution to journalArticle

Cikhardt, J, Klír, D, Kubeš, P, Kravárik, J, Řezáč, K, Šíla, O, Shishlov, AV & Labetsky, AY 2013, 'The construction of the fast resistive bolometer for a SXR measurement on the GIT-12 facility', Acta Polytechnica, vol. 53, no. 2, pp. 134-137.
Cikhardt J, Klír D, Kubeš P, Kravárik J, Řezáč K, Šíla O et al. The construction of the fast resistive bolometer for a SXR measurement on the GIT-12 facility. Acta Polytechnica. 2013;53(2):134-137.
Cikhardt, Jakub ; Klír, Daniel ; Kubeš, Pavel ; Kravárik, Josef ; Řezáč, Karel ; Šíla, Ondřej ; Shishlov, Alexander V. ; Labetsky, Alexey Yu. / The construction of the fast resistive bolometer for a SXR measurement on the GIT-12 facility. In: Acta Polytechnica. 2013 ; Vol. 53, No. 2. pp. 134-137.
@article{ddbac9c8343142f1878256b5f45cb9c2,
title = "The construction of the fast resistive bolometer for a SXR measurement on the GIT-12 facility",
abstract = "A lot of kinds of instruments are used for the SXR measurement at pulsed power facilities, but most of them are difficult to calibrate absolutely. For the determination of the energy of SXR radiated by the discharge on Z-pinches, it is possible to use the bolometer which can be calibrated analytically. The bolometer can be constructed with the sufficient sensitivity and, at the same time, with the time resolution in the order of nanoseconds. This bolometer was designed and constructed for the measurement on the 5MA facility GIT-12 at the Institute of High Current Electronics (IHCE) of the Siberian Branch Russian Academy of Sciences in Tomsk. The experiments on GIT-12 with the neon and deuterium gas-puff load were diagnosed by the copper bolometer with the time resolution of 4 ns and the sensitivity of 12Vcm2 J-1.",
keywords = "Bolometers, Plasma diagnostics, X-rays, Z-pinches",
author = "Jakub Cikhardt and Daniel Kl{\'i}r and Pavel Kubeš and Josef Krav{\'a}rik and Karel Řez{\'a}č and Ondřej Š{\'i}la and Shishlov, {Alexander V.} and Labetsky, {Alexey Yu}",
year = "2013",
language = "English",
volume = "53",
pages = "134--137",
journal = "Acta Polytechnica",
issn = "1210-2709",
publisher = "Czech Technical University, Publishing House",
number = "2",

}

TY - JOUR

T1 - The construction of the fast resistive bolometer for a SXR measurement on the GIT-12 facility

AU - Cikhardt, Jakub

AU - Klír, Daniel

AU - Kubeš, Pavel

AU - Kravárik, Josef

AU - Řezáč, Karel

AU - Šíla, Ondřej

AU - Shishlov, Alexander V.

AU - Labetsky, Alexey Yu

PY - 2013

Y1 - 2013

N2 - A lot of kinds of instruments are used for the SXR measurement at pulsed power facilities, but most of them are difficult to calibrate absolutely. For the determination of the energy of SXR radiated by the discharge on Z-pinches, it is possible to use the bolometer which can be calibrated analytically. The bolometer can be constructed with the sufficient sensitivity and, at the same time, with the time resolution in the order of nanoseconds. This bolometer was designed and constructed for the measurement on the 5MA facility GIT-12 at the Institute of High Current Electronics (IHCE) of the Siberian Branch Russian Academy of Sciences in Tomsk. The experiments on GIT-12 with the neon and deuterium gas-puff load were diagnosed by the copper bolometer with the time resolution of 4 ns and the sensitivity of 12Vcm2 J-1.

AB - A lot of kinds of instruments are used for the SXR measurement at pulsed power facilities, but most of them are difficult to calibrate absolutely. For the determination of the energy of SXR radiated by the discharge on Z-pinches, it is possible to use the bolometer which can be calibrated analytically. The bolometer can be constructed with the sufficient sensitivity and, at the same time, with the time resolution in the order of nanoseconds. This bolometer was designed and constructed for the measurement on the 5MA facility GIT-12 at the Institute of High Current Electronics (IHCE) of the Siberian Branch Russian Academy of Sciences in Tomsk. The experiments on GIT-12 with the neon and deuterium gas-puff load were diagnosed by the copper bolometer with the time resolution of 4 ns and the sensitivity of 12Vcm2 J-1.

KW - Bolometers

KW - Plasma diagnostics

KW - X-rays

KW - Z-pinches

UR - http://www.scopus.com/inward/record.url?scp=84877298405&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84877298405&partnerID=8YFLogxK

M3 - Article

AN - SCOPUS:84877298405

VL - 53

SP - 134

EP - 137

JO - Acta Polytechnica

JF - Acta Polytechnica

SN - 1210-2709

IS - 2

ER -