RANGE- TIME-OF-FLIGHT SPECTROMETER FOR MULTIPLY CHARGED IONS.

V. M. Bystritskii, A. G. Sterligov, Yu N. Usov

Research output: Contribution to journalArticlepeer-review

Abstract

In order to investigate the acceleration of bunches of multiply charged ions in a high-current electron beam, a spectrometer has been designed for multiply charged ions in which the ion energy is determined according to the time of flight over a base 30-cm long, while the charge state is determined from the magnitude of the range in a stack of absorptive coils. The time resolution of the spectrometer for operation with a 6LOR-2M oscilloscope is 3 nsec while the energy resolution is 30% at the 20-MeV level (scaled for hydrogen ions).

Original languageEnglish
Pages (from-to)643-645
Number of pages3
JournalInstrum Exp Tech
Volume19
Issue number3 pt 1
Publication statusPublished - 1976

ASJC Scopus subject areas

  • Instrumentation

Fingerprint Dive into the research topics of 'RANGE- TIME-OF-FLIGHT SPECTROMETER FOR MULTIPLY CHARGED IONS.'. Together they form a unique fingerprint.