Gradiometer Based on Two Single-Hole SQUID's

A. G. Likhachev, V. N. Polushkin, S. V. Uchaikin, B. V. Vasiliev

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

A gradiometer based on two ceramic superconducting quantum interferometers operating at liquid nitrogen temperature has been developed. The minimum gradient sensitivity is 3.1 x 10 -12T/mHz1/2 in a white noise region. The minimum damping coefficient of the external disturbing factors is K = 32.

Original languageEnglish
Pages (from-to)1034-1037
Number of pages4
JournalIEEE Transactions on Instrumentation and Measurement
Volume39
Issue number6
DOIs
Publication statusPublished - 1 Jan 1990
Externally publishedYes

Fingerprint

gradiometers
SQUIDs
Liquid nitrogen
White noise
Interferometers
Damping
white noise
liquid nitrogen
damping
ceramics
gradients
Temperature
sensitivity
coefficients
temperature

ASJC Scopus subject areas

  • Instrumentation
  • Electrical and Electronic Engineering

Cite this

Gradiometer Based on Two Single-Hole SQUID's. / Likhachev, A. G.; Polushkin, V. N.; Uchaikin, S. V.; Vasiliev, B. V.

In: IEEE Transactions on Instrumentation and Measurement, Vol. 39, No. 6, 01.01.1990, p. 1034-1037.

Research output: Contribution to journalArticle

Likhachev, A. G. ; Polushkin, V. N. ; Uchaikin, S. V. ; Vasiliev, B. V. / Gradiometer Based on Two Single-Hole SQUID's. In: IEEE Transactions on Instrumentation and Measurement. 1990 ; Vol. 39, No. 6. pp. 1034-1037.
@article{d65724854f3f4704a1f0afd62cea3003,
title = "Gradiometer Based on Two Single-Hole SQUID's",
abstract = "A gradiometer based on two ceramic superconducting quantum interferometers operating at liquid nitrogen temperature has been developed. The minimum gradient sensitivity is 3.1 x 10 -12T/mHz1/2 in a white noise region. The minimum damping coefficient of the external disturbing factors is K = 32.",
author = "Likhachev, {A. G.} and Polushkin, {V. N.} and Uchaikin, {S. V.} and Vasiliev, {B. V.}",
year = "1990",
month = "1",
day = "1",
doi = "10.1109/19.65821",
language = "English",
volume = "39",
pages = "1034--1037",
journal = "IEEE Transactions on Instrumentation and Measurement",
issn = "0018-9456",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
number = "6",

}

TY - JOUR

T1 - Gradiometer Based on Two Single-Hole SQUID's

AU - Likhachev, A. G.

AU - Polushkin, V. N.

AU - Uchaikin, S. V.

AU - Vasiliev, B. V.

PY - 1990/1/1

Y1 - 1990/1/1

N2 - A gradiometer based on two ceramic superconducting quantum interferometers operating at liquid nitrogen temperature has been developed. The minimum gradient sensitivity is 3.1 x 10 -12T/mHz1/2 in a white noise region. The minimum damping coefficient of the external disturbing factors is K = 32.

AB - A gradiometer based on two ceramic superconducting quantum interferometers operating at liquid nitrogen temperature has been developed. The minimum gradient sensitivity is 3.1 x 10 -12T/mHz1/2 in a white noise region. The minimum damping coefficient of the external disturbing factors is K = 32.

UR - http://www.scopus.com/inward/record.url?scp=0025539002&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0025539002&partnerID=8YFLogxK

U2 - 10.1109/19.65821

DO - 10.1109/19.65821

M3 - Article

VL - 39

SP - 1034

EP - 1037

JO - IEEE Transactions on Instrumentation and Measurement

JF - IEEE Transactions on Instrumentation and Measurement

SN - 0018-9456

IS - 6

ER -