Feasibility of clinical electron beam formation using polymer materials produced by fused deposition modeling

Irina Miloichikova, Angelina Bulavskaya, Yury Cherepennikov, Boris Gavrikov, Elisabetta Gargioni, Dmitrij Belousov, Sergei Stuchebrov

Research output: Contribution to journalArticle

Abstract

The main challenge in electron external beam radiation therapy with clinical accelerators is the absence of integrated systems to form irregular fields. The current approach to provide conformal irradiation is to use additional metallic shaping blocks, with inefficient and expensive workflows. This work presents a simple method to form therapeutic electron fields using 3D printed samples. These samples are manufactured by fused deposition modeling, which can affect crucial properties, such as material homogeneity, due to the presence of residual air-filled cavities. The applicability of this method was therefore investigated with a set of experiments and Monte Carlo simulations aimed at determining the electron depth dose distribution in polymer materials. The results show that therapeutic electron beams with energies 6–20 MeV can be effectively absorbed using these polymeric samples. The model developed in this study provides a way to assess the dose distribution in such materials and to calculate the appropriate thickness of polymer samples for therapeutic electron beam formation. It is shown that for total absorption of 6 MeV electron beams the material thickness should be at least 4 cm, while this value should be at least 8 cm for 12 MeV and 11 cm for 20 MeV, respectively. The results can be used to further develop 3D printing procedures for medical electron beam profile formation, allowing the creation of a collimator or absorber with patient-specific configuration using rapid prototyping systems, thus contributing to improve the accuracy of dose delivery in electron radiotherapy within a short manufacturing time.

Original languageEnglish
Pages (from-to)188-194
Number of pages7
JournalPhysica Medica
Volume64
DOIs
Publication statusPublished - 1 Aug 2019

Fingerprint

Polymers
electron beams
Electrons
polymers
dosage
radiation therapy
electrons
rapid prototyping
collimators
printing
homogeneity
absorbers
delivery
accelerators
Radiotherapy
manufacturing
cavities
irradiation
Workflow
air

Keywords

  • ABS and HIPS plastics
  • Electron beam
  • Fused deposition modeling
  • Radiation therapy

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging
  • Physics and Astronomy(all)

Cite this

Feasibility of clinical electron beam formation using polymer materials produced by fused deposition modeling. / Miloichikova, Irina; Bulavskaya, Angelina; Cherepennikov, Yury; Gavrikov, Boris; Gargioni, Elisabetta; Belousov, Dmitrij; Stuchebrov, Sergei.

In: Physica Medica, Vol. 64, 01.08.2019, p. 188-194.

Research output: Contribution to journalArticle

Miloichikova, Irina ; Bulavskaya, Angelina ; Cherepennikov, Yury ; Gavrikov, Boris ; Gargioni, Elisabetta ; Belousov, Dmitrij ; Stuchebrov, Sergei. / Feasibility of clinical electron beam formation using polymer materials produced by fused deposition modeling. In: Physica Medica. 2019 ; Vol. 64. pp. 188-194.
@article{af3c4a408a904641bb744e26f5ec8748,
title = "Feasibility of clinical electron beam formation using polymer materials produced by fused deposition modeling",
abstract = "The main challenge in electron external beam radiation therapy with clinical accelerators is the absence of integrated systems to form irregular fields. The current approach to provide conformal irradiation is to use additional metallic shaping blocks, with inefficient and expensive workflows. This work presents a simple method to form therapeutic electron fields using 3D printed samples. These samples are manufactured by fused deposition modeling, which can affect crucial properties, such as material homogeneity, due to the presence of residual air-filled cavities. The applicability of this method was therefore investigated with a set of experiments and Monte Carlo simulations aimed at determining the electron depth dose distribution in polymer materials. The results show that therapeutic electron beams with energies 6–20 MeV can be effectively absorbed using these polymeric samples. The model developed in this study provides a way to assess the dose distribution in such materials and to calculate the appropriate thickness of polymer samples for therapeutic electron beam formation. It is shown that for total absorption of 6 MeV electron beams the material thickness should be at least 4 cm, while this value should be at least 8 cm for 12 MeV and 11 cm for 20 MeV, respectively. The results can be used to further develop 3D printing procedures for medical electron beam profile formation, allowing the creation of a collimator or absorber with patient-specific configuration using rapid prototyping systems, thus contributing to improve the accuracy of dose delivery in electron radiotherapy within a short manufacturing time.",
keywords = "ABS and HIPS plastics, Electron beam, Fused deposition modeling, Radiation therapy",
author = "Irina Miloichikova and Angelina Bulavskaya and Yury Cherepennikov and Boris Gavrikov and Elisabetta Gargioni and Dmitrij Belousov and Sergei Stuchebrov",
year = "2019",
month = "8",
day = "1",
doi = "10.1016/j.ejmp.2019.07.014",
language = "English",
volume = "64",
pages = "188--194",
journal = "Physica Medica",
issn = "1120-1797",
publisher = "Associazione Italiana di Fisica Medica",

}

TY - JOUR

T1 - Feasibility of clinical electron beam formation using polymer materials produced by fused deposition modeling

AU - Miloichikova, Irina

AU - Bulavskaya, Angelina

AU - Cherepennikov, Yury

AU - Gavrikov, Boris

AU - Gargioni, Elisabetta

AU - Belousov, Dmitrij

AU - Stuchebrov, Sergei

PY - 2019/8/1

Y1 - 2019/8/1

N2 - The main challenge in electron external beam radiation therapy with clinical accelerators is the absence of integrated systems to form irregular fields. The current approach to provide conformal irradiation is to use additional metallic shaping blocks, with inefficient and expensive workflows. This work presents a simple method to form therapeutic electron fields using 3D printed samples. These samples are manufactured by fused deposition modeling, which can affect crucial properties, such as material homogeneity, due to the presence of residual air-filled cavities. The applicability of this method was therefore investigated with a set of experiments and Monte Carlo simulations aimed at determining the electron depth dose distribution in polymer materials. The results show that therapeutic electron beams with energies 6–20 MeV can be effectively absorbed using these polymeric samples. The model developed in this study provides a way to assess the dose distribution in such materials and to calculate the appropriate thickness of polymer samples for therapeutic electron beam formation. It is shown that for total absorption of 6 MeV electron beams the material thickness should be at least 4 cm, while this value should be at least 8 cm for 12 MeV and 11 cm for 20 MeV, respectively. The results can be used to further develop 3D printing procedures for medical electron beam profile formation, allowing the creation of a collimator or absorber with patient-specific configuration using rapid prototyping systems, thus contributing to improve the accuracy of dose delivery in electron radiotherapy within a short manufacturing time.

AB - The main challenge in electron external beam radiation therapy with clinical accelerators is the absence of integrated systems to form irregular fields. The current approach to provide conformal irradiation is to use additional metallic shaping blocks, with inefficient and expensive workflows. This work presents a simple method to form therapeutic electron fields using 3D printed samples. These samples are manufactured by fused deposition modeling, which can affect crucial properties, such as material homogeneity, due to the presence of residual air-filled cavities. The applicability of this method was therefore investigated with a set of experiments and Monte Carlo simulations aimed at determining the electron depth dose distribution in polymer materials. The results show that therapeutic electron beams with energies 6–20 MeV can be effectively absorbed using these polymeric samples. The model developed in this study provides a way to assess the dose distribution in such materials and to calculate the appropriate thickness of polymer samples for therapeutic electron beam formation. It is shown that for total absorption of 6 MeV electron beams the material thickness should be at least 4 cm, while this value should be at least 8 cm for 12 MeV and 11 cm for 20 MeV, respectively. The results can be used to further develop 3D printing procedures for medical electron beam profile formation, allowing the creation of a collimator or absorber with patient-specific configuration using rapid prototyping systems, thus contributing to improve the accuracy of dose delivery in electron radiotherapy within a short manufacturing time.

KW - ABS and HIPS plastics

KW - Electron beam

KW - Fused deposition modeling

KW - Radiation therapy

UR - http://www.scopus.com/inward/record.url?scp=85069690145&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85069690145&partnerID=8YFLogxK

U2 - 10.1016/j.ejmp.2019.07.014

DO - 10.1016/j.ejmp.2019.07.014

M3 - Article

VL - 64

SP - 188

EP - 194

JO - Physica Medica

JF - Physica Medica

SN - 1120-1797

ER -