Abstract
The influence of the adsorbed mass of molybdenum on the width of eluation profiles of generators and the patterns of molybdenum distribution in the amount of chromatographic columns by scanning them on the germanium-gallium detector using collimating device are studied. The boundary conditions under which the maximum value of 99mTc yield from generators Ye = 1 are defined. After scanning the columns, it was found out that the degree of filling the columns with molybdenum Q and the value of its maximum adsorption depend naturally on the total weight of the adsorbed mass. In order to achieve the condition Ye = 1 the value of Q should be at least 85%.
Original language | English |
---|---|
Title of host publication | Physics of Cancer |
Subtitle of host publication | Interdisciplinary Problems and Clinical Applications, PC 2016: Proceedings of the International Conference on Physics of Cancer: Interdisciplinary Problems and Clinical Applications 2016 |
Publisher | American Institute of Physics Inc. |
Volume | 1760 |
ISBN (Electronic) | 9780735414181 |
DOIs | |
Publication status | Published - 2 Aug 2016 |
Event | International Conference on Physics of Cancer: Interdisciplinary Problems and Clinical Applications 2016, PC 2016 - Tomsk, Russian Federation Duration: 22 Mar 2016 → 25 Mar 2016 |
Conference
Conference | International Conference on Physics of Cancer: Interdisciplinary Problems and Clinical Applications 2016, PC 2016 |
---|---|
Country | Russian Federation |
City | Tomsk |
Period | 22.3.16 → 25.3.16 |
ASJC Scopus subject areas
- Physics and Astronomy(all)