Direct Voltammetric Determination of Total Iron with a Gold Microelectrode Ensemble

E. A. Zakharova, E. E. Elesova, G. N. Noskova, M. Lu, R. G. Compton

    Research output: Contribution to journalArticle

    10 Citations (Scopus)

    Abstract

    The behavior of the Fe(II)/(III) redox system at a Au microelectrode ensemble (Au-MEE) based on a solid composite by means of direct and cyclic voltammetric analysis (VA) is reported. With a simple electrode activation and sample preparation, the influence of dissolved organic substances was eliminated, providing highly sensitive results. The analytical signal was based on the maximum cathodic current (I) of the first derivative (dI/dE), and iron determination within the 0.002-0.04mgL -1 range was studied. A sensitive LOD (3σ) value of 0.7μgL -1 for total iron concentration was calculated; total iron determination in different waters was shown.

    Original languageEnglish
    Pages (from-to)2061-2069
    Number of pages9
    JournalElectroanalysis
    Volume24
    Issue number11
    DOIs
    Publication statusPublished - 1 Nov 2012

    Keywords

    • Gold microelectrode ensemble
    • Iron
    • Microelectrodes
    • Voltammetric determination

    ASJC Scopus subject areas

    • Analytical Chemistry
    • Electrochemistry

    Fingerprint Dive into the research topics of 'Direct Voltammetric Determination of Total Iron with a Gold Microelectrode Ensemble'. Together they form a unique fingerprint.