Determination of Benzocaine Using HPLC and FIA with Amperometric Detection on a Carbon Paste Electrode

Hana Dejmkova, Veronika Vokalova, Jiri Zima, Jiri Barek

Research output: Contribution to journalArticle

9 Citations (Scopus)

Abstract

A new method for benzocaine determination employing FIA and HPLC with electrochemical detection on a carbon paste electrode was developed. The optimum conditions for the determination were found. Carrier solution for FIA consisted of B-R buffer pH4 (80% methanol, v/v) and used flow rate was 1.0mLmin-1. Mobile phase for HPLC consisted of B-R buffer pH4 (75% methanol, v/v) with flow rate 0.4mLmin-1. Working potential of +1.2V was employed. Practical applicability of the methods was tested on the determination of benzocaine in selected pharmaceuticals. The results were in agreement with results obtained using spectrophotometric detection and with one exception also with the content declared by the manufacturer.

Original languageEnglish
Pages (from-to)662-666
Number of pages5
JournalElectroanalysis
Volume23
Issue number3
DOIs
Publication statusPublished - 1 Mar 2011
Externally publishedYes

Fingerprint

Benzocaine
Ointments
Methanol
Buffers
Carbon
Flow rate
Electrodes
Drug products
Pharmaceutical Preparations

Keywords

  • Amperometric detection
  • Carbon paste electrode
  • FIA
  • HPLC
  • Pharmaceuticals

ASJC Scopus subject areas

  • Analytical Chemistry
  • Electrochemistry

Cite this

Determination of Benzocaine Using HPLC and FIA with Amperometric Detection on a Carbon Paste Electrode. / Dejmkova, Hana; Vokalova, Veronika; Zima, Jiri; Barek, Jiri.

In: Electroanalysis, Vol. 23, No. 3, 01.03.2011, p. 662-666.

Research output: Contribution to journalArticle

Dejmkova, Hana ; Vokalova, Veronika ; Zima, Jiri ; Barek, Jiri. / Determination of Benzocaine Using HPLC and FIA with Amperometric Detection on a Carbon Paste Electrode. In: Electroanalysis. 2011 ; Vol. 23, No. 3. pp. 662-666.
@article{f2db7a9a6e4248e5bd9d2188871aa9ff,
title = "Determination of Benzocaine Using HPLC and FIA with Amperometric Detection on a Carbon Paste Electrode",
abstract = "A new method for benzocaine determination employing FIA and HPLC with electrochemical detection on a carbon paste electrode was developed. The optimum conditions for the determination were found. Carrier solution for FIA consisted of B-R buffer pH4 (80{\%} methanol, v/v) and used flow rate was 1.0mLmin-1. Mobile phase for HPLC consisted of B-R buffer pH4 (75{\%} methanol, v/v) with flow rate 0.4mLmin-1. Working potential of +1.2V was employed. Practical applicability of the methods was tested on the determination of benzocaine in selected pharmaceuticals. The results were in agreement with results obtained using spectrophotometric detection and with one exception also with the content declared by the manufacturer.",
keywords = "Amperometric detection, Carbon paste electrode, FIA, HPLC, Pharmaceuticals",
author = "Hana Dejmkova and Veronika Vokalova and Jiri Zima and Jiri Barek",
year = "2011",
month = "3",
day = "1",
doi = "10.1002/elan.201000646",
language = "English",
volume = "23",
pages = "662--666",
journal = "Macromolecular Chemistry and Physics",
issn = "1040-0397",
publisher = "Wiley-VCH Verlag",
number = "3",

}

TY - JOUR

T1 - Determination of Benzocaine Using HPLC and FIA with Amperometric Detection on a Carbon Paste Electrode

AU - Dejmkova, Hana

AU - Vokalova, Veronika

AU - Zima, Jiri

AU - Barek, Jiri

PY - 2011/3/1

Y1 - 2011/3/1

N2 - A new method for benzocaine determination employing FIA and HPLC with electrochemical detection on a carbon paste electrode was developed. The optimum conditions for the determination were found. Carrier solution for FIA consisted of B-R buffer pH4 (80% methanol, v/v) and used flow rate was 1.0mLmin-1. Mobile phase for HPLC consisted of B-R buffer pH4 (75% methanol, v/v) with flow rate 0.4mLmin-1. Working potential of +1.2V was employed. Practical applicability of the methods was tested on the determination of benzocaine in selected pharmaceuticals. The results were in agreement with results obtained using spectrophotometric detection and with one exception also with the content declared by the manufacturer.

AB - A new method for benzocaine determination employing FIA and HPLC with electrochemical detection on a carbon paste electrode was developed. The optimum conditions for the determination were found. Carrier solution for FIA consisted of B-R buffer pH4 (80% methanol, v/v) and used flow rate was 1.0mLmin-1. Mobile phase for HPLC consisted of B-R buffer pH4 (75% methanol, v/v) with flow rate 0.4mLmin-1. Working potential of +1.2V was employed. Practical applicability of the methods was tested on the determination of benzocaine in selected pharmaceuticals. The results were in agreement with results obtained using spectrophotometric detection and with one exception also with the content declared by the manufacturer.

KW - Amperometric detection

KW - Carbon paste electrode

KW - FIA

KW - HPLC

KW - Pharmaceuticals

UR - http://www.scopus.com/inward/record.url?scp=79954625849&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=79954625849&partnerID=8YFLogxK

U2 - 10.1002/elan.201000646

DO - 10.1002/elan.201000646

M3 - Article

AN - SCOPUS:79954625849

VL - 23

SP - 662

EP - 666

JO - Macromolecular Chemistry and Physics

JF - Macromolecular Chemistry and Physics

SN - 1040-0397

IS - 3

ER -