Compact 150 kV electron accelerator

G. M. Kassirov, F. G. Sekisov, N. N. Shumilova

Research output: Contribution to journalArticle

Abstract

The design and testing results were described of electron accelerator based on pulse transformer with high-voltage unit vacuum insulation. The output pulse of the transformer was up to 150 kV. The current pulses of up to 5 mA and ≈200 ns length at the electron energy in output beam of ≈60 keV were obtained. The electron beam energy was sufficient for colouring alkaline-haloid crystals, while the irradiation of fluoroplastic for several minutes caused structural transformations in it. When testing accelerator mockup the break-downs from winding last turn to camera were observed. These break-downs were probably connected with the organic substance presence in operating zone. The service life and pulse recurrence rate were limited by K15-4-40 condensator serviceablity in shaping line.

Original languageEnglish
Pages (from-to)171-172
Number of pages2
JournalPribory i Tekhnika Eksperimenta
Issue number2
Publication statusPublished - 1 Mar 1991

ASJC Scopus subject areas

  • Instrumentation

Fingerprint Dive into the research topics of 'Compact 150 kV electron accelerator'. Together they form a unique fingerprint.

  • Cite this

    Kassirov, G. M., Sekisov, F. G., & Shumilova, N. N. (1991). Compact 150 kV electron accelerator. Pribory i Tekhnika Eksperimenta, (2), 171-172.