Accuracy of the Garmin 920 XT HRM to perform HRV analysis

Johan Cassirame, Romain Vanhaesebrouck, Simon Chevrolat, Laurent Mourot

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

Heart rate variability (HRV) analysis is widely used to investigate autonomous cardiac drive. This method requires periodogram measurement, which can be obtained by an electrocardiogram (ECG) or from a heart rate monitor (HRM), e.g. the Garmin 920 XT device. The purpose of this investigation was to assess the accuracy of RR time series measurements from a Garmin 920 XT HRM as compared to a standard ECG, and to verify whether the measurements thus obtained are suitable for HRV analysis. RR time series were collected simultaneously with an ECG (Powerlab system, AD Instruments, Castell Hill, Australia) and a Garmin XT 920 in 11 healthy subjects during three conditions, namely in the supine position, the standing position and during moderate exercise. In a first step, we compared RR time series obtained with both tools using the Bland and Altman method to obtain the limits of agreement in all three conditions. In a second step, we compared the results of HRV analysis between the ECG RR time series and Garmin 920 XT series. Results show that the accuracy of this system is in accordance with the literature in terms of the limits of agreement. In the supine position, bias was 0.01, − 2.24, + 2.26 ms; in the standing position, − 0.01, − 3.12, + 3.11 ms respectively, and during exercise, − 0.01, − 4.43 and + 4.40 ms. Regarding HRV analysis, we did not find any difference for HRV analysis in the supine position, but the standing and exercise conditions both showed small modifications.

Original languageEnglish
Pages (from-to)831-839
Number of pages9
JournalAustralasian Physical and Engineering Sciences in Medicine
Volume40
Issue number4
DOIs
Publication statusPublished - 1 Dec 2017
Externally publishedYes

Keywords

  • Accuracy
  • Garmin 920XT
  • HRV
  • Validity

ASJC Scopus subject areas

  • Biophysics
  • Biomedical Engineering
  • Radiology Nuclear Medicine and imaging

Fingerprint Dive into the research topics of 'Accuracy of the Garmin 920 XT HRM to perform HRV analysis'. Together they form a unique fingerprint.

  • Cite this